T. Ching, J. K. Chin, J. Tan, C. Y. Ong, K. S. Raditia, S. Y. Lau, J. Quah, X. Khoo
Undergraduate capstone, SUTD, with Changi General Hospital and Eastern Health Alliance
Featured in The Straits Times. [Read the article]
Pleural effusion is a buildup of fluid around the lungs. To treat it, a tube is placed in the chest to drain the fluid. Draining too fast or too much can be dangerous, so nurses and doctors have to monitor the process closely. Our hospital partners wanted a device that could control the drainage automatically, so patients would be safer and staff would be freed up.
Instead of designing a custom pump, we built the device around parts every hospital already has: a standard syringe and a 3-way tap. In the pull phase, the syringe draws fluid from the chest. In the push phase, the tap switches and the syringe pushes the fluid into a collection bag. Repeating this cycle can drain any volume, at a controlled rate. A stepper motor and lead screw move the plunger precisely, and a second motor with feedback control switches the tap between its two positions. Because the syringe and tap are standard parts, the disposables are cheap and easy to replace.
A medical device has to fail safely. We added sensors that stop the drainage if air bubbles appear in the line, which can mean a loose or dislodged tube, and that detect blockages from blood clots or debris. A proximity sensor checks that the syringe, tap, and tubing are fitted correctly before the program starts, and that drainage continues safely even if the device is knocked over.
The final prototype is smaller than a traditional drain box, drained fluid at up to 2.13 mL/s within the safe range, and had no leaks across 24 trials. We delivered it under budget. The project was featured in The Straits Times, and the hospital filed a patent on the design.