T. Ching, J. Vasudevan, H.Y. Tan, C.T. Lim, J. Fernandez, Y.-C. Toh, M. Hashimoto
HardwareX, 10(2021) e00202. [Paper Link]
Almost every microfluidic and organ-on-chip experiment needs a pump. But pumps suited for microfluidics are bulky and expensive, often over $1,000 each, and you can't modify them. Every experiment has different needs, like flow rate, number of channels, or space inside an incubator. I wanted researchers to be able to tailor the pump to their experiment, not the other way around.
I designed a kit to build your own peristaltic pump from 3D-printed parts and off-the-shelf electronics, for about $50. A peristaltic pump moves fluid by rolling over a soft tube, squeezing liquid forward with each turn. I used a tiny geared stepper motor that an Arduino can drive directly, which kept the pump small and cheap. The smallest version is 20 × 50 × 28 mm, and the pumps cover flow rates from 0.02 to 727 µL/min.
Instead of a finished product, I designed the pump as a kit. Every part mounts on a breadboard with holes spaced 10 mm apart, like studs on Lego bricks. This makes it easy to swap parts for different tubing sizes, rearrange the layout for different experiments, or replace a worn part without rebuilding the whole pump. It also means new modules can be added later. To show this, I built a flow rate calibration tool that uses two infrared sensors to time the liquid as it passes, and it plugged right into the existing setup.
The pumps ran inside a cell incubator for seven days straight while perfusing a cell culture, with no drop in performance, and parts showed little wear after 60 days of continuous use. I used these pumps to run the perfusion cultures in my vascular work. All design files and code are open source.
All design files and code are open source, and a detailed build protocol can be found in this paper.