C., Juan P. F., T. Ching, C. Wu, C. Y. Chionh, N. Suranga, S. Foong
IEEE CIS & RAM 2015 [Paper Link]
Featured in The Straits Times. [Read the article]
Undergraduate UROP, with Changi General Hospital.
After a central venous catheter is removed, the wound can start bleeding heavily in the first few hours. The site is covered in thick bandages and often blankets, so bleeding can go unnoticed between nurse checks, which happen every 15 to 30 minutes. In rare cases, this can be fatal. Our clinical partners wanted a way to catch bleeding the moment it starts.
BWard is a small, coin-sized sensor that sits on top of the dressing. It works in two steps. First, it checks whether the dressing is wet, using an electrical moisture sensor and infrared light. If it finds fluid, it then shines green light into the dressing. Blood absorbs green light strongly, while water, urine, and serum barely do. If the reflected light drops, the fluid is blood. This two-step approach means the device only raises an alarm for bleeding, not for sweat or other fluids.
I designed and fabricated the BWard's housing. It had to do three jobs at once. First, it had to block outside light, since the device detects blood by measuring reflected light and any stray light would cause false readings. Second, it had to keep blood and other fluids away from the electronics. Third, it had to be small and light enough to sit comfortably on a bandaged wound. The result is a two-part, 4 cm housing. The outer cover faces the dressing, shields the device, and carries the moisture-sensing circuit. The inner core holds the main circuit board and battery. The whole device weighs about 52 g.
We tested BWard on hundreds of dressings with artificial blood at different volumes and dilutions, and it reliably told blood apart from serum even at low concentrations. In a first test with real blood from hemodialysis patients, it detected bleeding in every case, with under 1.1 mL of blood.